It allows to take the measurements of:
- earth resistance using auxiliary electrodes,
- earth resistance using auxiliary electrodes and clamp (for measurements of multiple earth)
- earth resistance using double clamps (for measurement of earthing when it is impossible to use auxiliary electrodes),
- ground resistivity (Wenner method),
- measurement of continuity of equipotential bondings and protective conductors (meeting the requirements of IEC 60364 – 6-61:2000 section 6.12.2) with auto-zero function – with current 200 mA.
Additionally:
- measurement of resistance of auxiliary electrodes RS and RH,
- measurement of interference voltage,
- measurement of interference frequency,
- measurement in the presence of interference voltage in the power network with frequency 50 Hz, 60 Hz
- selection of maximum measuring voltage (25 V and 50 V)
- introducing the distance between the electrodes for the resistivity in metres (m),
- memory of 990 measurements (10 banks of 99 cells each),
- calibration of clamp used,
- real time clock (RTC),
- data transmission to the computer (USB, wireless),
- indication of battery state.
Electric security:
- type of insulation: double, according to EN 61010 – 1 and IEC 61557
- measurement category: CAT IV 300 V acc. to EN 61010 – 1
- protection class acc. to EN 60529: IP54
Rated operational conditions:
- operation temperature: -10…+50°C
- storage temperature: -20…+70°C
- humidity: 20…80%
Other technical data:
- LCD display: graphic, backlighted
- interface: USB, wireless
- number of measurements carried out of set of batteries: > 500
- warranty: 24 months
Measurement of interference voltage
0…100 V | 1 V | ±(2% m.v + 3 digits) |
- singnalling overvoltage 24 V or 40 V rms
- measurement for DC and AC 45…65 Hz,
Measurement of earth resistance (method 3- and 4-pole)
measurement range to IEC61557 – 5: 0,30 Ω…19,9 kΩ
0,00…19,99 Ω | 0,01 Ω | ±(2% m.v. + 2 digits) |
20,0…199,9 Ω | 0,1 Ω |
200…1999 Ω | 1 Ω |
2,0k…9,99 kΩ | 0,01 kΩ | ±(5% m.v. + 4 digits) |
10,0k…19,9 kΩ | 0,1 kΩ |
- measurement current: under short circuit >200 mA,
- voltage on open terminals: selectable <25 V AC or <50 V AC,
- frequency of measurement current: 125 (for networks 50 Hz) or 150 Hz (for 60 Hz),
Measurement of continuity of equipotential bondings and protective conductors (Rcont)
measurement range to IEC61557 – 5: 0,24 Ω…19,9 kΩ
0,00…19,99 Ω | 0,01 Ω | ±(2% m.v. + 2 digits) |
20,0…199,9 Ω | 0,1 Ω |
200…1999 Ω | 1 Ω |
2,0k…9,99 kΩ | 0,01 kΩ | ±(5% m.v. + 2 digits) |
10,0k…19,9 kΩ | 0,1 kΩ |
- voltage on open terminals: <24 Vrms but >4 Vrms,
- measurement current: under short circuit >200 mA,
- frequency of measurement current: 125 (for networks 50 Hz) or 150 Hz (for networks 60 Hz), possible choice of measurement frequency in menu,
- auto-zero function measurement leads
Measurement of resistance of auxiliary electrodes RH and RS
0…999 Ω | 1 Ω | ±(5% (RS+RE+RH) + 8 digits) |
1,00k…9,99 kΩ | 0,01 kΩ |
10,0k…19,9 kΩ | 0,1 kΩ |
Measurement of multiple earth resistance with using the clamp and auxiliary electrodes (3p + clamp)
0,00…19,99 Ω | 0,01 Ω | ±(8% m.v. + 3 cyfry) |
20,0…199,9 Ω | 0,1 Ω |
200…1999 Ω | 1 Ω |
- voltage on open terminals: selectable <25 V AC or <50 V AC,
- measurement current: under short circuit > 200 mA,
- frequency of measurement current: 125 (for networks 50 Hz) or 150 Hz (for 60 Hz),
Measurement of multiple earth resistance with using double clamps
0,00..19,99 Ω | 0,01 Ω | ±(10% m.v. + 3 digits) |
20,0…149,9 Ω | 0,1 Ω | ±(20% m.v. + 3 digits) |
- frequency of measurement current 125 (for networks 50 Hz) or 150 Hz (for networks 60 Hz)
Measurement of ground resistivity
Measurement method: Wennera, ρ=2πLRE
0,0…199,9 Ωm | 0,1 Ωm | depending on measurement accuracy RE with 4p method, but not less than ±1 digit |
200…1999 Ωm | 1 Ωm |
2,00k…19,99 kΩm | 0,01 kΩm |
20,0k…99,9 kΩm | 0,1 kΩm |
100k…999 kΩm | 1 kΩm |
L – distance between probes: 1…50 m